In Part 1 we showed that a biosensor’s claims decide its framework: wellness, medical device or in vitro diagnostic, and which class within it.
This part covers what comes next, with a focus on electrochemical biosensors: enzymatic sensors under the skin, glucose test strips, sweat patches and the immunosensor cartridges used at the point of care. Once the claim has fixed the framework, it also fixes how much evidence you need before you can sell. In the EU that evidence is a clinical evaluation or a performance evaluation. In the US it is the testing behind a 510(k) or a De Novo. The question is the same in every case: does the evidence support exactly what you say the product does?
1. The principle: the evidence follows the claim
In the EU, proportionality is written into the regulations
Annex XIV of the MDR says that the clinical evaluation’s “depth and extent shall be proportionate and appropriate to the nature, classification, intended purpose and risks of the device in question, as well as to the manufacturer’s claims in respect of the device.”
The IVDR sets a similar proportionality rule in Annex XIII, based on the device’s risks, risk class, performance and intended purpose, and MDCG 2022-2 puts it plainly: “Each indication and claimed clinical benefit specified in the intended purpose should be assessed and have the appropriate supporting clinical evidence.”
It is the manufacturer who must “specify and justify the level of clinical evidence necessary” (MDR Article 61(1); IVDR Article 56(1)). The notified body checks that justification and, under Annex IX of the MDR, assesses “to what extent specific claims are supported by specific pre-clinical and clinical data” for any characteristic claimed as innovative and for new indications.
MDR and IVDR ask for different kinds of evidence
This is why the framework decision in Part 1 matters so much for electrochemical sensors, which can fall on either side of the line.
- Under the MDR, you carry out a clinical evaluation (Article 61, Annex XIV). Its building blocks are clinical data: clinical investigations, scientific literature, and post-market data.
- Under the IVDR, you carry out a performance evaluation (Article 56, Annex XIII) built on three pillars: scientific validity (is the analyte associated with the condition?), analytical performance (does the device measure it correctly?) and clinical performance (do the results correlate with the clinical condition in the target population?).
A glucose sensor under the skin and a glucose test strip measure the same molecule with similar chemistry. The first needs a clinical evaluation under the MDR; the second needs a performance evaluation under the IVDR. They are different studies, with different rules.
Equivalence is narrower than most teams expect
Under the MDR you can rely on clinical data from another device only if it is equivalent in its technical, biological and clinical characteristics (Annex XIV, Section 3). For an electrochemical sensor, the list is demanding:
- Technical: similar design, “similar principles of operation and critical performance requirements” and similar “software algorithms”.
- Biological: “the same materials or substances in contact with the same human tissues or body fluids for a similar kind and duration of contact”. A different membrane, enzyme coating or electrode material in contact with interstitial fluid already puts this in question.
- Clinical: same condition, same site in the body, a similar population and the same kind of user.
For devices that are neither implantable nor class III, the MDR does not require a contract with the other manufacturer (MDCG 2020-5, MDCG 2023-7), but you still need “sufficient levels of access” to the data. Without it, “equivalence claims cannot be made for the purpose of conformity assessment.” We covered which data sources are acceptable in our article on similar-device data.
Most performance standards are not harmonised
ISO 15197, the standard for blood glucose meters for self-testing, is not harmonised under the IVDR. Most standards for physiological sensors are not harmonised under the MDR either. You can and should use them as the state of the art, but they do not give a presumption of conformity under Article 8, and you have to justify them in the technical documentation.
Two standards that matter for electrochemical sensors are harmonised:
- EN ISO 10993-1:2025 (biological evaluation) under the MDR, since 17 June 2026 — the first time any edition of 10993-1 has been on the MDR list;
- EN ISO 20916:2024 (clinical performance studies for IVDs) under the IVDR.
In the US, clinical data in a 510(k) is the exception
A 510(k) requires substantial equivalence to a predicate, and the regulation asks for clinical data only “if deemed necessary” (21 CFR 807.100(b)). The FDA’s draft guidance on clinical data in 510(k)s (September 2023, still a draft) lists when it may be needed, including different indications, different technology, or performance that bench testing alone cannot show. One of its examples is directly relevant: an IVD intended for point-of-care use whose predicate is a laboratory test may need clinical data, because the clinical settings and patient populations differ.
For CGMs and brain injury tests, however, the FDA has written the evidence into the regulation itself as special controls, as the next sections show.
Optical wearables (ECG, SpO₂, blood pressure) are outside the scope of this part. Their framework was covered in Part 1, and the FDA’s draft guidances on pulse oximeters (January 2025) and cuffless blood pressure (January 2026) set out its current thinking on their evidence.
2. Continuous monitors under the skin: glucose, ketones, lactate
EU
A CGM is a medical device under the MDR (MDCG 2024-11, see Part 1). It needs a clinical evaluation and, because the filament sits in interstitial fluid, biocompatibility data. EN ISO 10993-1:2025, now harmonised, is the reference for that evaluation. EN ISO 10993-10 (skin sensitisation) and 10993-23 (irritation), relevant to the adhesive patch, are harmonised too.
A sensor worn for 10 to 15 days is not an implantable device under the MDR, which requires at least 30 days in place (Article 2(5)). So the requirement to perform clinical investigations for implantable and class III devices (Article 61(4)) does not apply to it by default. But without an equivalent device whose data it can access, a new CGM has to generate its own clinical data: Annex XIV requires new data wherever gaps in the evidence remain. The investigation will be invasive, so it needs national authorisation (Article 70(7)(b)).
Do not reach for ISO 15197 here: it covers capillary blood meters, not interstitial glucose. For CGM performance metrics, the FDA recognises CLSI POCT05.
US
The FDA has set numerical acceptance criteria for integrated CGMs in the regulation itself (21 CFR 862.1355). In the adult clinical study, against a laboratory reference method traceable to a higher order, the lower bound of the one-sided 95% confidence interval must exceed the following:
| CGM reading | Within | Must exceed |
|---|---|---|
| Below 70 mg/dL | ±15 mg/dL | 85% |
| 70–180 mg/dL | ±15% | 70% |
| Above 180 mg/dL | ±15% | 80% |
| Whole measuring range | ±20% | 87% |
There are also rules against serious errors: when the CGM reads below 70 mg/dL, no reference value may be above 180 mg/dL, and the reverse. Other special controls cover interferents (“endogenous substances and metabolites, foods, dietary supplements, and medications” — a real issue for enzymatic electrodes), a usability study, and consistent performance “throughout the sensor wear period”.
Three points from 2026 are worth knowing:
- OTC CGMs meet the same special controls as prescription ones. Over-the-counter use did not lower the bar.
- The paediatric OTC clearance (June 2026) reused existing data. According to the public 510(k) summary, the manufacturer reprocessed data from a previously completed clinical study with the new algorithm, instead of running a new paediatric study. For days 11 to 15 of wear it relied on a justification, not on direct data in children.
- The first glucose and ketone monitor (De Novo, August 2026) was supported by six clinical studies with more than 600 participants. The new regulation it created applies the same glucose thresholds as for CGMs. For ketones, accuracy must be “adequate to ensure clinically acceptable agreement, as determined by FDA”, plus two rules against gross errors, with no fixed percentages. Anyone who files after it will follow these special controls.
Lactate and other analytes. There is no regulation with numerical thresholds for continuous lactate. Without a predicate, the first manufacturer of a new type of moderate-risk device usually goes through De Novo, and its request must include a draft of the special controls that later products will have to meet (21 CFR 860.220).
Microneedles. In the US they cannot be wellness products (Part 1), so the evidence route is the device route from the start.
3. Glucose test strips and meters
EU
A finger-prick brings your study under the full study rules. Under Article 58(1) of the IVDR, a clinical performance study that involves surgically invasive sampling done only for the study must follow the requirements of Articles 58 to 77, including an application to the competent authority. MDCG 2025-5 (June 2025) lists capillary blood sampling as surgically invasive, so a finger-prick taken just for your study is enough. Whether you then have to wait for authorisation depends on the risk: if the sampling is not a major clinical risk, you can generally start once the application has been validated, unless national law says otherwise (Article 66(7)(a)). Studies on leftover samples follow a lighter regime.
Legacy meters have until the end of 2027. Under the old Directive 98/79/EC, blood glucose meters for self-testing were on List B, so they needed a notified body certificate. Under Article 110(3a) and (3c) of the IVDR, as amended by Regulation (EU) 2024/1860, a meter with that certificate can stay on the market until 31 December 2027, but only if the manufacturer met a series of conditions, including:
- a quality management system under Article 10(8) in place by 26 May 2025;
- a formal application to an IVDR notified body by 26 May 2025;
- a written agreement with that notified body by 26 September 2025.
The later date often quoted for class C, 31 December 2028, applies to devices that were self-declared under the Directive. It does not apply to meters, which already had a certificate. The Commission’s reform proposal of December 2025 does not change these dates.
What the IVDR asks for. Under the IVDR, a meter for self-testing is a class C IVD (Rule 4(a), see Part 1). The performance evaluation covers the three pillars, and self-testing adds specific requirements:
- performance “obtained by laypersons” (Annex I, Section 9.4);
- a design suited to “the skills and the means available to the intended user”, reducing as far as possible the risk of error in interpreting results and, where feasible, warning the user if “the device has failed to provide a valid result” (Annex I, Section 19);
- for the notified body, “studies carried out with intended users” and the labelling and instructions (Annex IX, Section 5.1).
ISO 15197 sets the accuracy and user-performance requirements for these systems, but it is not harmonised under the IVDR. A new edition is in development.
EUDAMED. Since 28 May 2026, the first four EUDAMED modules are mandatory, including the registration of economic operators and devices.
US
Blood glucose meters are class II devices under 21 CFR 862.1345 and reach the market through a 510(k). The FDA does not recognise ISO 15197. It has two final guidances of September 2020, one for each use, and they set different accuracy targets:
- Over-the-counter use: 95% of results within ±15% of the comparator across the whole measuring range, and 99% within ±20%. The user evaluation needs at least 350 subjects, and at least 10% of them should never have used a meter.
- Prescription point-of-care use (hospitals and clinics): 95% of results within ±12% above 75 mg/dL, or ±12 mg/dL below it, and 98% within ±15% or ±15 mg/dL. At least 350 patients for each sample type.
A meter that meets the ISO 15197 criteria will not necessarily meet the FDA ones, so plan the US studies against these guidances from the start. Since 2 February 2026, the FDA’s quality system rule (QMSR) also incorporates ISO 13485:2016.
4. Sweat patches and new analytes
There is no official EU guidance on the evidence for sweat, tear or breath sensors. The evidence follows the framework, so the borderline decision in Part 1 also decides which kind of evidence you need.
The reference problem. Electrochemical sweat sensors often measure analytes, or matrices, for which no reference method exists. The IVDR addresses this directly in Annex XIII: for “novel markers or other markers without available certified reference materials or reference measurement procedures, it may not be possible to demonstrate trueness.” Where there are no comparative methods either, “a clinical performance study comparing performance of the novel device to the current clinical standard practice is required.” MDCG 2022-2 adds that a “high degree of novelty (such as new analytes and/or new intended purposes)” requires new data and a scientific rationale.
Scientific validity comes first. Before you show that the patch measures sweat lactate or sweat glucose correctly, you have to show that the value in sweat is associated with the condition you claim. For many sweat analytes, that association is the weakest part of the file.
A patch that collects and analyses. If the device is applied to the skin and contains an integrated reagent that analyses the sweat, one could infer that MDR Article 1(7) applies: the whole product would be governed by the MDR, with the IVD part meeting the IVDR requirements. No official document says this for sweat patches. The inference comes from two similar examples in MDCG 2024-11, a CGM with an integrated capillary blood measurement and a saliva device with integrated reagents.
For this family, the most expensive mistake is to design the study before the framework is settled.
5. Immunosensors and point-of-care cartridges
EU: performance evaluation, step by step
Annex I of the IVDR lists the analytical characteristics to address: analytical sensitivity and specificity, trueness, precision, limits of detection and quantitation, measuring range, linearity, cut-off, interference and cross-reactions. It also lists the clinical ones: diagnostic sensitivity and specificity, predictive values and likelihood ratios. Annex XIII then sets the rule for the evidence: “as a general rule”, analytical performance “shall always be demonstrated on the basis of analytical performance studies”, and clinical performance studies “shall be performed unless due justification is provided” for relying on other sources.
For near-patient tests, the performance must be shown “in relevant environments (for example, patient home, emergency units, ambulances)” (Annex I, Section 9.4), and the notified body examines studies with the intended users (Annex IX, Section 5.1). MDCG 2022-2 adds that a reader with its own measuring function also needs scientific validity data, “unless due justification is given.”
For class D devices, where an EU reference laboratory has been designated, the notified body must ask it to verify the claimed performance by laboratory testing (Article 48(5)).
US: what the FDA reviews
The FDA’s public decision summaries show what a 510(k) for a point-of-care cartridge contains. For a high-sensitivity troponin I cartridge (January 2025), the summary covers precision at three point-of-care sites, linearity, interference, detection limits, cut-offs, method comparison, and studies on haematocrit, altitude and vibration. The clinical study was run at 28 sites with 3,585 patients presenting to the emergency department, adjudicated by blinded cardiologists and emergency physicians.
For brain injury markers, the special controls of 21 CFR 866.5830 set out the clinical study: patients with suspected mild traumatic brain injury (GCS 13–15), a head CT (or another appropriate diagnostic standard) for every eligible subject, each scan read blind by at least two radiologists, operators and settings representative of the intended use, and development samples kept out of the validation. The regulation sets no minimum sensitivity or negative predictive value; it requires you to measure and report them. A whole-blood cartridge cleared in March 2024 reported a clinical sensitivity of 96.5% and a specificity of 40.3% in 970 evaluable patients.
CLIA waiver. A test intended for sites without a laboratory needs evidence that untrained operators get the same results. The FDA guidance (February 2020) recommends at least three sites and at least nine untrained operators in total, plus “flex studies” that stress the test’s operational limits. The dual 510(k) and CLIA waiver pathway lets one set of studies serve both purposes, and the FDA describes it as in many instances “the least burdensome and fastest approach.”
What the claim decides. For immunosensors the claim decides the study population, the comparator and the setting. “Aids in assessing the need for a CT scan” needs a different study from “detects brain injury”.
6. Implantable biosensors
This is where the MDR is strictest. As Part 1 explained, implantable and class III devices must undergo clinical investigations (Article 61(4)), with narrow exceptions, and the 2026 list of well-established technologies does not include sensors.
What it means for evidence:
- Non-clinical data alone is not an option. The Article 61(10) route cannot be used for class III or implantable devices (MDCG 2020-6).
- Class III implantable devices go through the clinical evaluation consultation procedure (Article 54): an expert panel can comment on the notified body’s assessment of the clinical evidence.
- The post-market work is heavier. The PMCF evaluation report and the summary of safety and clinical performance must be updated “at least annually” (Article 61(11)).
- Long-term biocompatibility — enzyme layers, membranes and electrode materials in contact with tissue for months — is assessed against the now harmonised EN ISO 10993-1:2025.
For an implantable sensor, the clinical development plan, from first-in-human to pivotal investigation and PMCF, is part of the clinical evaluation plan that Annex XIV requires, and a notified body will expect to see it early.
7. What the reform would change in the evidence
The Commission’s proposal (COM(2025) 1023), introduced in Part 1, would change several of the rules above. In the MDR it would:
- Widen clinical data to include studies on the device itself published in scientific literature, “but not necessarily peer-reviewed” (recital 9). Post-market clinical follow-up data already counts today.
- Make non-clinical evidence easier to use. Article 61(10) would add in vitro, ex vivo and in silico testing, computational modelling and simulation, and “the data available for the generic device group”. It would still be an exception that must be justified, and still not available where Article 61(4) requires clinical investigations.
- Relax equivalence. “The same or similar” materials and clinical condition instead of “the same”, and no contract needed to avoid a clinical investigation under Article 61(5). The technical criteria would not change.
- Narrow the duty to run clinical investigations. Article 61(4) would apply to “implantable class IIb devices and class III devices” instead of all implantable devices, so an implantable class IIa sensor would no longer be covered by it.
- Drop the separate PMCF evaluation report. PMCF findings would go straight into the clinical evaluation report.
- Create an adapted pathway for breakthrough and orphan devices, allowing certification “even if based on limited clinical data” where post-market follow-up fills the gap.
- Define “well-established technology devices”. By its criteria (“little evolution in indications”, “a long history on the Union market”), a novel implantable sensor is unlikely to qualify. That is our reading, not an official position.
In the IVDR it would allow clinical performance to rest on data from an equivalent device for the first time, accept literature that is not peer-reviewed, mention computational modelling and in silico testing, and let manufacturers of class C and D devices ask an expert panel for early advice on their clinical strategy. Clinical performance studies would still be required “unless due justification is provided.”
None of this is law. As of mid-November 2026, the procedure (2025/0404(COD)) is still awaiting a committee decision in Parliament, with an indicative plenary date of 18 January 2027. The Council’s June progress report states that more work is needed to reach a general approach; delegations broadly support the changes on evidence, but several “remain cautious regarding the cumulative impact of some of the simplifications.” A file submitted today is assessed against the current MDR and IVDR.
8. Before you design a study
- Write the claim first, then design the study. Every word in the intended purpose — population, setting, user, “monitors”, “detects”, “rules out” — becomes a study requirement.
- Settle the framework before the protocol. A clinical investigation under the MDR and a clinical performance study under the IVDR are different studies.
- Check equivalence honestly. A different enzyme layer, membrane, electrode material or algorithm usually means your own data.
- Find your reference method early. For new analytes and matrices such as sweat, the absence of a reference method can decide the whole study design.
- Check which standards are harmonised. ISO 15197 is not; EN ISO 10993-1:2025 and EN ISO 20916:2024 are.
- Plan for both markets at once. FDA special controls can set numerical thresholds that a European file does not require, and the FDA does not recognise some standards that Europe relies on.
Check the claim before you design the study
Most evidence problems begin with a claim that says more than the data can support. RegAffairs Hub offers a fixed-scope claims and intended purpose audit: we read everything you say about your product, the way a notified body or the FDA would, and tell you which framework, class and evidence those claims point to, before you commit to a study.
Get in touchSources
- EU legislation: MDR, Regulation (EU) 2017/745 · IVDR, Regulation (EU) 2017/746 · Implementing Decision (EU) 2026/1231 (harmonised standards, MDR) · Implementing Decision (EU) 2024/2625 (EN ISO 20916:2024, IVDR) · Harmonised standards, medical devices · Harmonised standards, IVD
- MDCG guidance: MDCG 2020-5, equivalence (April 2020) · MDCG 2020-6, sufficient clinical evidence (April 2020) · MDCG 2023-7, exemptions and access to data (December 2023) · MDCG 2022-2, clinical evidence for IVDs (January 2022) · MDCG 2024-11, qualification of IVDs (October 2024) · MDCG 2025-5, performance studies Q&A (June 2025)
- EU reform: COM(2025) 1023 · COM(2025) 1023, Annex · Legislative Observatory, 2025/0404(COD) · Council progress report, ST 9801/26 (June 2026)
- FDA: 21 CFR 807.100 (substantial equivalence) · 21 CFR 860.220 (De Novo content) · Draft guidance, clinical data in 510(k) submissions (Sept 2023) · 21 CFR 862.1355 (iCGM) · FDA recognised standard CLSI POCT05 (Rec. 7-307) · 510(k) summary K260935 (paediatric OTC CGM) · De Novo DEN250049 (glucose and ketone monitor) · FDA press release, first glucose and ketone monitor (25 Aug 2026) · Decision summary K240984 (hs-troponin I) · 21 CFR 866.5830 (brain trauma assessment test) · Decision summary K234143 (whole-blood TBI cartridge) · CLIA waiver guidance (Feb 2020) · Dual 510(k) and CLIA waiver guidance (Feb 2020)